Siemens CH5-2 Convex Ultrasound Probe 08648086 Image Artifacts Repair ~ 40891

Probe initializes and produces images but has persistent imaging artifacts, nonuniform grayscale response, and inconsistent Color Doppler performance. Sold AS-IS for repair, refurbishment, or parts; full functionality is not guaranteed. Not suitable for clinical use without professional repair and testing.

More details


$216.00

40891

Availability: This product is no longer in stock.

Product details

Siemens CH5-2 Convex Ultrasound Probe 08648086 Image Artifacts AS-IS Repair

Condition: Pre-Owned | Good Cosmetic Condition | Tested – Image Artifacts Present | AS-IS – Repair Recommended


Device / Instrument Profile
Manufacturer: Siemens Medical Solutions
Model: CH5-2   P/N: 08648086      S/N: 82435416
Probe Type: Curved Array / Convex Ultrasound Transducer
Frequency Range: 2–5 MHz     DOM: 2008
Protection Rating: IPX8
Cable Length: Approx. 8 ft
Test System: Siemens SONOLINE G40


Item Overview
Siemens CH5-2 curved-array ultrasound transducer designed for compatible Siemens ultrasound systems. The probe initializes and produces ultrasound images; however, persistent imaging abnormalities were observed during testing and repair is recommended.


Condition & Inspection
Probe Housing: Good cosmetic condition with normal signs of previous use. No visible cracks or significant structural damage observed.

Acoustic Lens: Appears intact with no visible cuts, tears, or surface separation.

Cable Assembly: Approx. 8-ft cable appears intact with no visible cuts, tears, or exposed wiring.

Connector: Connector housing, contacts, and locking mechanism appear intact.


Functionality / Testing
Initialization: Successfully recognized by the Siemens SONOLINE G40 and loaded the appropriate imaging preset.

Air Scan: Persistent image irregularities and nonuniformity were observed during testing.

Phantom Imaging: Grayscale images were generated; however, localized artifacts and image-uniformity issues remained present.

Body Imaging: Probe produced recognizable abdominal images, but image quality was affected by nonuniform signal response and acoustic artifacts.

Color Doppler: Color-flow signals were displayed; however, visualization appeared patchy and inconsistent during testing.

Assessment: Findings are consistent with suspected transducer-element degradation, element dropout, or another internal imaging defect.

Testing Result: Probe initializes and produces grayscale and Color Doppler images; however, persistent imaging artifacts are present and diagnostic image quality is not considered satisfactory.


Included

  • 1 × Siemens CH5-2 Curved Array Ultrasound Transducer
  • Attached Approx. 8-ft Cable
  • Integrated Siemens Connector Assembly

Not Included
Siemens SONOLINE G40 Ultrasound System, Carrying Case, User Manual, Accessories, Consumables, or Additional Equipment.


Buyer Advisory
The actual probe pictured is the unit supplied.

This transducer is being sold AS-IS for repair, refurbishment, parts harvest, evaluation, or non-clinical applications.

The probe initializes and generates ultrasound images; however, persistent image artifacts, nonuniform grayscale response, and inconsistent Color Doppler performance were observed during testing.

Professional evaluation and repair are recommended before any diagnostic use.

Please review all testing photographs carefully, including air-scan, phantom, body-scan, and Doppler images
.

 


Stock ID: 40891 (P.O 2027) (WH104 B2)