For orders, please call: 952-232-5970

$0Checkout 0 Item(s)

MENU

    

Medical Positioning (MPI) S...

Medical Positioning (MPI) Stress 1736 Echo Bed/Patient Positioning Table ~ 40016

Item specifics

Condition
For parts or not working
Seller Notes
“Pre‑owned MPI Stress 1736 Echo Bed (S/N 0805490, USA) powered patient positioning table. Tested functional for Trendelenburg, Reverse Trendelenburg, lateral tilt, lowering; lift actuator verified with alternate controller."
Model
1736
MPN
1736

More details


$823.20

40016

1 item in stock

Warning: Last item in stock!

Add to cart

Place an offer

Product details

Medical Positioning (MPI) Stress 1736 Echo Bed / Powered Patient Positioning Table – Side Tilt – Trendelenburg – UP/DOWN

Device Information

  • Manufacturer: Medical Positioning, Inc. (MPI)

  • Item: Stress Echo Bed / Powered Patient Positioning Table

  • MPI Item #: 1736

  • Serial Number: 0805490

  • Country of Manufacture: USA

  • Electrical Rating: 120 VAC • 60 Hz • 1.6 A

Item Overview

Pre‑owned MPI Stress 1736 Echo Bed / powered patient positioning table in good cosmetic and functional condition.

The table was tested using the included MPI wired hand control. All powered functions—Trendelenburg, Reverse Trendelenburg, left/right lateral tilt, and lowering—operated normally. However, the UP (raise) button on the included hand control is non‑functional.

To confirm the lift actuator, a compatible MPI hand control from another table was temporarily connected. With the compatible controller, the table raised normally, verifying that the lift actuator and drive system are fully operational. The issue appears isolated to the included hand control. Replacement is recommended.

The table includes dual mechanical drop‑out sections, two side‑rail mounting clamps, an accessory receiver clamp, multiple accessory mounting sockets, an integrated patient restraint strap, and lockable casters.

Grey medical‑grade upholstery is in good overall condition. Minor tears are present on the side panels; primary patient contact surfaces show no significant damage.

Cosmetic Condition

  • Good pre‑owned cosmetic condition

  • Grey medical‑grade upholstery

  • Main patient surface in good condition

  • Minor tears on side upholstery panels

  • Normal cosmetic wear from previous clinical use

  • Frame and base structurally sound

  • Lockable casters in good condition

Functional Testing

  • Unit powers on normally

  • Included hand control tested

  • UP (raise) button on included hand control is unresponsive

  • DOWN function operates normally

  • Powered Trendelenburg verified

  • Powered Reverse Trendelenburg verified

  • Powered left/right lateral tilt verified

  • Lift actuator verified using compatible MPI hand control

  • Table raises normally with compatible controller

  • Dual mechanical drop‑out sections functional

  • Latching mechanisms operate correctly

  • Table stable throughout tested range

  • Locking caster system verified

Drive System

  • SKF electric linear actuator system

  • Multiple actuators for height, tilt, and Trendelenburg functions

  • Lift actuator confirmed operational

  • Included hand control has defective UP button and should be replaced

Accessories Included

  • MPI wired hand control (UP button defective)

  • Integrated patient restraint strap

  • Two side‑rail mounting clamps

  • Accessory receiver clamp

  • Multiple accessory mounting sockets

  • Lockable caster base

Not Included

  • Side rails

  • IV pole

  • Instrument tray

  • Additional positioning accessories

  • User manual

Quick Condition Snapshot

✔ Good cosmetic condition ✔ Main upholstery in good condition ✔ Lift actuator verified functional ✔ Trendelenburg / Reverse Trendelenburg operational ✔ Left/right lateral tilt operational ✔ Dual drop‑out sections functional ✔ Side‑rail clamps included ✔ Patient restraint strap included ✔ Accessory receiver clamp & mounting sockets included ✔ Included hand control has non‑functional UP button ✔ Replacement hand control recommended

Buyer Advisory

The included MPI hand control has a non‑functional UP (raise) button. All other powered functions operate normally. Testing with a compatible MPI hand control confirmed the lift actuator and drive system are fully functional. Buyers should plan to repair or replace the included hand control for full operation.

Please review all listing photographs carefully. Buyers are responsible for verifying compatibility with their intended application and performing any recommended inspection or preventive maintenance before clinical use.

Shipping

Freight shipping only. Professionally palletized, secured, and protected for safe transport. Liftgate and residential delivery available upon request (additional charge).


Stock ID: 40016 (P.O. 2026) (WH104) 

Purchase Terms & Buyer Protections

Condition & Use Notice

Condition: Overall, cosmetic, and functional condition are described in the listing above; please review all images carefully.
Functionality: Sold as described in the listing, or untested if stated. Biomed evaluation is recommended where applicable.
Included: Only items shown in photographs or specifically stated in the listing are included.
Buyer Responsibility: Buyer is responsible for verifying compatibility, suitability, and obtaining inspection, calibration, maintenance, or service as appropriate prior to use.