Siemens ACUSON 14L5 SP Linear Ultrasound Probe S2000 Image Dropout AS-IS ~ 40770

Item specifics

Condition
For parts or not working
Seller Notes
“Siemens ACUSON 14L5 SP linear-array ultrasound probe, P/N 10041226, S/N 64410542, tested on a Siemens ACUSON S2000 system.Significant image dropout on the left side of the scan field persisted during phantom scanning;
Model
14L5SP
MPN
14L5SP

More details


$55.20

40770

1 item in stock

Warning: Last item in stock!

Add to cart

Place an offer

Product details

Siemens ACUSON 14L5 SP Linear Ultrasound Probe – Tested – Image Dropout – AS-IS

Item Overview

Pre-owned Siemens ACUSON 14L5 SP linear-array ultrasound transducer, tested on a Siemens ACUSON S2000 ultrasound system.

Significant image dropout was observed on the left side of the scan field during testing. Approximately half of the image field remained functional, with grayscale and Color Doppler activity demonstrated in the functioning region.

Probe failed image-quality testing and requires repair. Sold strictly AS-IS for parts, repair, or refurbishment.

Device Information

Manufacturer: Siemens Medical Solutions USA, Inc.
Model: 14L5 SP (14L5SP)
Part Number: 10041226
S/N: 64410542
DOM: November 2016
Probe Type: Linear Array Ultrasound Transducer
Cable Length: Approx. 8 ft
Connector: Pinless Locking Type
IP Rating: IPX8
Made In: Korea

Condition & Inspection

Cosmetic: Good overall pre-owned condition. Probe housing, handle, and acoustic lens appear intact with no visible tears.

Cable & Connector:

  • Cable insulation intact with no visible cuts.
  • Strain relief intact.
  • Connector housing and locking mechanism appear intact.
  • Copper contact surface shows oxidation/patina.

Functional Check:

  • Tested on a Siemens ACUSON S2000 ultrasound system.
  • Probe recognized and presets loaded successfully.
  • Significant image dropout present on the left side of the scan field.
  • Dropout persisted during phantom scanning.
  • Right carotid artery visualized in the functioning portion of the image.
  • Color Doppler demonstrated in the functioning region.

Assessment: Findings suggest possible transducer element failure; exact defect not determined.

Included

  • 1× Siemens ACUSON 14L5 SP Ultrasound Probe with Attached Cable

Not Included

  • Ultrasound System
  • Storage Case
  • Additional Accessories

Only the items shown and specifically listed are included.

Buyer Advisory

FOR PARTS OR NOT WORKING. AS-IS SALE.

This probe has a confirmed image-quality defect affecting a substantial portion of the scan field. It is being sold for parts, repair, refurbishment, or evaluation purposes only.

Image quality, clinical performance, and diagnostic suitability are not acceptable in its current condition. Repair, inspection, and complete performance testing are required before any clinical use.

Stock ID: 40770 (P.O.1200) (WH104 SCB2) 

Purchase Terms & Buyer Protections

Condition & Use Notice

Condition: Overall, cosmetic, and functional condition are described in the listing above; please review all images carefully.
Functionality: Sold as described in the listing, or untested if stated. Biomed evaluation is recommended where applicable.
Included: Only items shown in photographs or specifically stated in the listing are included.
Buyer Responsibility: Buyer is responsible for verifying compatibility, suitability, and obtaining inspection, calibration, maintenance, or service as appropriate prior to use.