GE M12L Linear Ultrasound P...
GE M12L Linear Ultrasound Probe Ref 2294512 | Tested – Image Dropout ~ 40641
Item specifics
Product details
GE M12L Linear Array Ultrasound Probe – LOGIQ 9 Tested – Image Dropout – Parts/Repair – P/N 2294512 – S/N 919702YM1
Item Overview
Pre‑owned GE M12L linear‑array ultrasound transducer, tested on a GE LOGIQ 9. Probe was recognized and produced an image, but major image dropout / severe image abnormality was present across air, phantom, and limited carotid testing. Offered AS‑IS / FOR PARTS OR REPAIR.
Device Information
Manufacturer: GE / GE Yokogawa Medical Systems Model: M12L Type: Linear Array Connector Model / P/N: 2294512 Serial Number: 919702YM1 Manufactured: Feb 2003 Made In: Japan CE: 0459 Cable Length: ~8 ft Connector: GE multi‑pin Condition: AS‑IS / Parts Repair
Physical Condition
Good overall cosmetic condition
Housing intact; no major cracks or dents
Acoustic lens smooth and intact
Cable intact; no obvious cuts/tears
Strain reliefs intact
Connector housing and contact arrays intact
Normal signs of previous use Review photos for exact cosmetic condition.
Functional Testing (LOGIQ 9)
Recognized as M12L
Air Scan: Major image dropout
Phantom Scan: Dropout/abnormality persisted
Carotid Scan: Anatomy partially visualized; severe dropout remained Actual test images included for buyer review.
No FirstCall or element‑level testing performed. Findings consistent with suspected nonfunctioning elements/channels, but exact defect not diagnosed.
Package Includes
1× GE M12L Linear Array Ultrasound Probe
Sale is for the pictured probe only.
Buyer Advisory
Sold AS‑IS / FOR PARTS OR REPAIR due to significant image dropout. Not patient‑ready. Buyer/facility must perform professional evaluation, repair, and full safety/performance validation before any clinical use.
Stock ID: 40641 (P.O. 1200) (05E)