Manufacturer | Verathon Medical |
Item: | Verathon BladderScan BVI9400 W/ Probe E570-0315, Batteries, Charger & PSU |
Model/Cat # | BVI 9400 / 0570-0190 |
Serial No. | B4307329 |
Made In: | USA |
Date of Manufacture: | 2017-01-18 |
Voltage: 5200mAh | 11.1V | 57Wh
Short description:
Verathon BladderScan BVI 9400 bladder volume instrument with probe, two rechargeable batteries, dual bay wireless charger, and AC adapter with power cord. Console powers on, self-test runs, probe responds with scans, and printer is functional (out of paper). Self-test shows EPROM communication failure — calibration/service is required.
Probe:
- Probe Reference: E570-0315
- Probe Serial Number: PE004198
- Probe DOM: 2021-02-17
Battery Packs (Qty: 2):
- Verathon Rechargeable Lithium-Ion
- P/N 0400-0066
- Rating: 11.1V, 5200mAh, 58Wh
- UL listed, Made in Korea
Battery Charger:
- Wireless Hub
- Ref 0570-0193
- Serial No# WO28576
- Verathon P/N 0400-0067
- Eljintek Model GPSU40B-5
- Input: 100–240V AC, 50/60Hz, 1.0A
- Output: 16–21V DC, 2.82–2.15A (set at 18V = 2.5A, 45W max)
- Made in Taiwan (Eljintek, Inc.)
Console:
Housing intact; light scratches/scuffs from use.
Protective LCD cover shows a small stress crack at the top edge (common).
LCD bright and responsive.
Handle and base feet intact.
Probe connector solid.
Built-in printer operational but no paper included.
Probe:
Housing intact, no dents or cracks.
Cable ~4 ft, no cuts/tears.
Strain relief secure.
Connector in good condition.
Battery Charger (Dual Bay):
Housing intact, no cracks.
LEDs functional (orange = charging, green = charged).
Battery Packs (x2):
Good cosmetic shape, no dents/cracks.
Charge accepted, but long-term health unknown.
Overnight charging recommended.
No warranty on used batteries.
Power Adapter:
Good condition, connector intact.
Includes power cord.
Self-Test Report:
Console S/N: 000307329 | Versions: U.I. 3.1.0.5/706, DSP 1.2/BVI9400 | Disk Space: 19482KB available | Microphone: Passed | DSP: Present and functional | Error: EPROM test failed
| (could not communicate with the battery) | Internal Temp: 25.25 °C | Calibration: Run ScanPoint required | Printer: Open/out of paper | Scan Counter: 012,409.
Note on EPROM Error: “EPROM test failed” means the system could not properly communicate with or verify its internal memory chip (EEPROM/EPROM). -it indicates a firmware or memory fault that can affect saving/calibration functions. The device still powers on, but reliability is uncertain until serviced.
Probe Testing: Mock scans on saline bottle produced volume readings.
Calibration: Not performed, system prompts for calibration.
Printer: Working, but no extra paper included.
Recommendation: Calibration/service required before clinical use.
- User manual
- Phantom
- Calibration kit
- Gel
- Mobility cart/stand
- Extra printer paper
- Any consumables or disposables
Package includes:
BladderScan BVI 9400 console (P/N 0570-0190)
Bladder scanner probe (Ref E570-0315)
2 × rechargeable battery packs (P/N 0400-0066, 11.1V, 5200mAh, 58Wh)
Dual bay wireless battery charger (P/N 0570-0193)
AC power adapter (Eljintek Model GPSU40B-5, P/N 0400-0067) W/ Power cord
*Items shown in photos are the items that will ship. If it is not shown in the photos, it is not included*
Sterilized : Non-sterile device
Manual : Not Included
Cleaned : Yes, at our facility
Shipping Weight : 30 lbs
Stock ID : 38523 (P.O 2017) (WH 555)
Comments:
Kindly enlarge the images and see them with zooming.
Concern or query: Ask a Question
Condition: USED, please refer to images for cosmetic condition.
- 1. Overall Condition:- As seen in images - Very Good
- 2. Physical Condition: Please review images before bidding - Good
- 3. Working Condition: Has been tested at our end | Calibration Required....Will receive item as pictured ……………(offered AS IS)
- 4. “As a whole; the item has not been tested technically or professionally at our end- unable to test --- being sold STRICTLY AS IS! Biomed check is strongly recommended before clinical use!”