For orders, please call: 952-232-5970

$0Checkout 0 Item(s)

MENU

    

Olympus BF‑P40 Flexible Fib...

Olympus BF‑P40 Flexible Fiberoptic Endoscopy Scope | Broken Image Fibers ~ 40052

Item specifics

Condition
For parts or not working
Seller Notes
“Pre‑owned Olympus BF‑Type P40 Flexible Fiberoptic Bronchoscope (S/N 1232747, Japan, CE 0197) offered FOR PARTS / REPAIR / AS‑IS. Scope articulates smoothly and focus adjustment works, but numerous broken fibers."
Model
BF-Type P40
MPN
BF-Type P40

More details


$363.20

40052

1 item in stock

Warning: Last item in stock!

Add to cart

Place an offer

Product details

Olympus BF-Type P40 Flexible Fiberoptic Bronchoscope – For Parts / Repair / AS-IS

Item Overview

Pre-owned Olympus BF-Type P40 Flexible Fiberoptic Bronchoscope being sold FOR PARTS / REPAIR / AS-IS due to numerous broken image fibers producing multiple black spots throughout the field of view. Overall cosmetic condition is good with normal signs of prior clinical use.

The bronchoscope was connected to a compatible Olympus endoscopy system (not included) to evaluate its optical image and basic mechanical operation. Articulation controls functioned as intended; however, the optical defects require professional repair before any clinical use.

Sale includes the bronchoscope only.


Device Information

Manufacturer: Olympus Medical Systems Corp.
Item: Flexible Fiberoptic Bronchoscope
Model: BF-Type P40 • Serial No.: 1232747
Made In: Japan • Reusable: Yes • Non-Sterile


Manufacturer Specifications

Optical System

  • Field of View: 120°
  • Depth of Field: 3–50 mm

Insertion Tube

  • Distal End Diameter: 4.9 mm
  • Insertion Tube Diameter: 5.0 mm
  • Working Length: 550 mm
  • Total Length: 840 mm

Instrument Channel

  • Channel Diameter: 2.2 mm
  • Minimum Visible Distance: 5 mm from Distal Tip

Bending Section

  • Up: 180°
  • Down: 130°

Condition Summary

Control Body

  • Good overall cosmetic condition
  • No major dents or structural cracks observed
  • Minor cosmetic wear consistent with previous clinical use
  • Control knobs rotate smoothly
  • Eyepiece/focus adjustment functions properly

Light Guide Connector & Universal Cord

  • Light guide connector in good cosmetic condition
  • No dents, cracks, or significant damage observed
  • Connector interfaces appear clean
  • Universal cord approximately 5 ft long
  • Flexible throughout
  • No cuts, tears, or exposed wiring observed
  • Strain reliefs intact

Insertion Tube & Distal Section

  • Exterior insertion tube in good cosmetic condition
  • Graduation markings remain legible
  • No obvious cuts, punctures, or major damage observed
  • Bending rubber appears intact
  • No tears or separation observed
  • Distal tip articulates properly with the control lever

Functional Evaluation

Mechanical Function

  • Articulation control lever operates smoothly
  • Distal tip deflects properly in the tested directions
  • Mechanical controls appear responsive
  • No excessive resistance noted during manipulation

Optical Evaluation

  • Connected to a compatible Olympus endoscopy system (not included)
  • Eyepiece and focus adjustment function properly
  • Numerous broken image fibers observed
  • Multiple black spots visible throughout the field of view, including against a white background
  • Optical condition is documented in the listing photographs

Overall Assessment

  • Mechanical functions appeared satisfactory during the limited evaluation performed
  • Significant optical defects require repair before clinical use
  • No leak test, pressure test, instrument channel patency verification, or complete performance evaluation was performed

Included

  • Olympus BF-Type P40 Flexible Fiberoptic Bronchoscope

Not Included

  • Endoscopy processor / video system
  • Light source
  • Storage case
  • Cleaning brushes
  • Biopsy forceps
  • User manual
  • Original packaging
  • Any accessories not shown

Cleaning / Sterility

Cleaned: Yes • Sterilized: No • Patient Ready: No • Reusable: Yes


Buyer Advisory

Please verify compatibility with your Olympus endoscopy equipment before purchase. The endoscopy processor and light source used during evaluation are not included and were utilized solely to demonstrate the bronchoscope's current condition. This device is not patient-ready and is being sold AS-IS for parts, repair, refurbishment, training, or research purposes due to the optical defects described above. Professional inspection, repair, and validation are recommended prior to any clinical application. Please review all listing photographs carefully, as they form an important part of the item description.

Stock ID: 40052 (P.O. 2026) (OFH C6)

Purchase Terms & Buyer Protections

Condition & Use Notice

Condition: Overall, cosmetic, and functional condition are described in the listing above; please review all images carefully.
Functionality: Sold as described in the listing, or untested if stated. Biomed evaluation is recommended where applicable.
Included: Only items shown in photographs or specifically stated in the listing are included.
Buyer Responsibility: Buyer is responsible for verifying compatibility, suitability, and obtaining inspection, calibration, maintenance, or service as appropriate prior to use.